Oxford Instruments has achieved the world's first 15 T integrated Cryogen-free dilution refrigerator and magnet system!
One single Pulse Tube Cooler with vibration isolation at 4 K and 70 K
Easy and reliable sample access : via only one O ring sealed vacuum space, no IVC
Load-lock mechanism for fast sample change
Split shields to allow full sample region access
Oxford instruments published an article in Cryogenics on the integration of superconducting magnets with cryogen-free dilution refrigerator systems.
Download the article here >>
Low temperature insert
Independent precool loop
Kelvinox dilution unit
2 x NW40 Line of site ports
1 x NW50 Line of site port
240 mm diameter x 240 mm length sample space
Diagnostic thermometry
Plain tail
Fully automatic gas handling
Software control
Turbo pump
Dry rotary pump
Condensing line compressor
Electrical isolation
3 T, 77 mm cold bore
8 T, 57 mm cold bore
12 T, 57 mm cold bore
15 T, 57 mm cold bore
0.1% homogeneity over 10 mm DSV
Superconducting switch available
Quench protection circuit
HTS current leads
Vector rotate magnets available on request
Remote rotary valve
Remote pumping station
LN
Optical tails
X-Ray or Neutron tails
Tails to suit room temperature bore magnet system
DC & RF customer wiring
RF heat sinking
Sample thermometry
Oxford Instruments has now delivered nine integrated dilution refrigerator and magnet cryofree systems. Six of those have already been installed. All systems have met or exceeded factory specifications. See details below.
Germany
November
VR2/0.2/0.2
7.3 mK
400 µW at 93.5 mK
USA
July 2009
0.3 T split pair
8.0 mK
200 µW at 83 mK
8T
12 mK
250 µW
April 2009
8 T
7 mK
400 µW
8.5 mK
240 µW
UK
March 2009
12 T
Switzerland
5 T
7.1 mK
235 µW
Australia
7.5 mK
300 µW
Canada
8.7 mK
3 T
6.8 mK